2.2 Specifications
2.2 Specifications
(1) Number of Flow Paths
6 (paths)
(2) Standard
throughput
3 mL/min (when using distilled water. Applicable range
varies with the kind of mobile phase to be used.
Note that, if momentary maximum flow exceeds 3 mL/min,
the degassing performance given in (3) below may not be
attained.)
(3) Degassing
Performance
UV absorbance change at wavelength 210 nm shall be
200 mAU or larger when the methanol (HPLC grade) in
which the air is dissolved up to saturation is delivered at
flow rate 1 mL/min (at liquid temperature 25
°
C, with cell
length 10 mm).
(4) Internal
Volume
About 480
μ
L/1 flow path (excluding tube)
(5) Vacuum
Level
6.7 kPa
(6) Maximum Withstand Pressure
0.2 MPa
Note that normal flow into the degassing unit shall be made
by means of suction.
A load of pressure to the flow path in the degassing unit
shall always be negative.
(7) Error Output Condition
(a) When the internal pressure of degassing chamber has
not reached 11 kPa at 10 minutes after start of the
vacuum pump drive.
(b) When vacuum level has exceeded 13 kPa.
2 - 2
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...