3.1 Turning on Power Supply
3.1.3 Auto
Zero
A/Z
(Auto zero)
ABS indication and output signal on the monitor screen can be
zeroed.
0.0 250 0.0000
1
D2
TIME
WL
ABS
PROG
LAMP [I]
Auto zero adjustable range: -0.2 to 2 AU
If this range is exceeded, the following message appears.
PRESS “CL” KEY TO CLEAR MESSAGE
A/Z OVER RANGE
NOTE
: Auto zero function is unusable during 2-
wavelength measurement.
3.1.4 Setting of Measuring Wavelength
A measuring wavelength is settable.
WL
(Wavelength)
The measuring wavelength setting screen appears.
250
D2 MODE (190 – 600), W MODE (380 – 900)
WAVELENGTH (190 – 900) nm
Press [numerical] keys (within [1][9][0] to [9][0][0]), and then the
[ENT] key.
* Settable wavelength range: 190 to 900 nm, enterable in
1-nm steps
3 - 8
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...