3 - 11
3.1.6 Manual
Operation
The operation is used for delivering liquid under the constant
condition without using a program.
(1) Isocratic
Mode
As an example: flow rate at 1.0 ml/min, upper pressure limit
at 20.0 MPa, and lower pressure limit at 0 MPa.
NOTE
:
The manual operation settings will be deleted when
the power supply is turned off. Conditions must be
set up each time when turning on the power supply.
NOTE
:
The pump noise will be changed with the operating
condition such as the flow rate, and the pressure.
Monitor screen
MANUAL
0
MANUAL: (FLOW=1, PRESS=2)
1
CONSTANT FLOW FLOW MAX MIN
1.000 20.0 0.0
Select Isocratic mode (FLOW)
1 ENT
CONSTANT FLOW FLOW PRESS [MN]
0.000 0.0 C 1
Enter FLOW rate, MAX and MIN pressure limit.
ENT
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...