4.4.3
WARNING
Inflammation or injury due to toxic, corrosive or
stimulative solvent!
When using a toxic, corrosive or stimulative solvent, be
careful not to incur inflammation or injury. For details of
the properties of each solvent and how to handle it, refer to
Material Safety Data Sheet (MSDS). Be sure to handle
each solvent properly.
(a) Wear proper protective means (e.g., safety goggles)
so that a solvent will not come into direct contact with
the skin, eye, etc.
(b) Ventilate the laboratory room adequately to prevent
accidental inhalation of harmful solvent vapor.
If there is contamination inside the flow cell, cleaning is required.
Items to be prepared for flow cell cleaning:
•
Injector (made of glass, having a capacity of about 10 mL)
•
Detergent
Cable connector for
thermostatic flow cell
control (option)
Clamp screw
Fig. 4-3 Check of Flow Cell
4 - 13
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...