<Wavelength setting screen (Ex indicated)>
Ex
Em
(200 - 850.0)
350 350
Press numeric keys 3 5 0 and then the ENT key.
(ii) An emission (Em) wavelength will not be set
here. Leave the present setting as it is.
<Wavelength setting screen (Em indicated)>
Ex Em
(
250 - 900.0
)
350
350
Press the ENT key.
(iii) The
1
st
monitor screen appears again.
(c) Set a time constant, a photomultiplier voltage and an
emission spectrum bandwidth.
(i) Press
the
UTLTY
SET UP
key, and the utility setting
screen appears.
<Utility setting screen>
UTILITY (OFFSET=1, TIME CONSTANT=2,
_2
USE PROG=3, PMT VOLT=4, OTHERS=5)
Press numeric key 2 and then the ENT key.
(ii) The time constant setting screen appears.
<Time constant setting screen>
TIME CONSTANT (0.01s=1, 0.02s=2, 0.05s=3,
7
0.1s=4, 0.5s=5, 1.0s=6, 2.0s=7)
Press numeric key 7 and then the ENT key.
(iii) The
1
st
monitor screen appears.
(iv) Press
UTLTY
SET UP
key. And the utility setting screen
reappears.
4 - 4
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...