2.2
(c) At occurrence of pressure sensor error
* Active status (error detected) can be checked from
the Chromato Data Station (CDS) or GUI controller.
(8) Material of Wetted Parts
TEFLON-AF (degassing membrane), glass fiber-reinforced
PPS (connector)
(9) Power
Supply
24 V DC
(10) Power Consumption
Max. 12 W (normally 9 W)
(11) Dimensions
53 (W)
×
297 (D)
×
100 (H) mm
(12) Operating Principle
The liquid to be degassed is flowed through a special
synthetic resin membrane placed in the vacuum chamber of
6-channel degassing unit. The molecules of gases
dissolved in liquid are small, and have a higher mobility
than the liquid and a high affinity to the resin membrane.
Therefore, the gases are selectively transmitted through the
membrane and discharged out of the degassing unit.
Thus, the gases are eliminated from the liquid.
2 - 3
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...