(7) Sample
loading
Deliver the sample into the sample loop.
The syringe discharges [vf + vi + va + vd : Feed
Injection Air Dead volume (inside
parameter)], and delivers a sample into the loop.
Buffer tube
vd :Dead volume (the instrument inside parameter)
vi :Injection volume (the sample net volume to the flow path)
vf :Feed volume (the wash fluid volume to deliver a sample to the loop)
va :Air volume (the air volume which sample is sandwiched.)
Syringe valve
Syringe
Wash pump
5
1
3
2
4
Plunger
wash
Sample
Rack
P
Needle
Injection valve
Injection port
5
4
3
2
1
6
Loop
vf
vf + vi
+ vd
+ va
Wash
fluid1
Wash
fluid2
From Pump To Column
Waste
solution
Waste
solution
Wash port
Degassing unit
Degassing
unit
3-way solenoid valve
2 - 33
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...