Press numeric keys 2 5 0 to 8 5 0 and then the ENT
key.
* Setting range
: 250 to 850 nm
* Initial (default) value : 250
* If setting will not be changed, press
V
key to
proceed to WL2 setting.
The cursor now moves to WL2.
(d) Now enter an end wavelength (maximum value) at
WL2.
<Scan range setting screen (indicate WL2)>
SCANNING WL RANG (WL1 WL2) nm FL (Em)
<300-900> 250
-
700
Press numeric keys 3 0 0 to 9 0 0 and then the ENT
key.
* Input value
: Value of less than 900 within
WL1 + 50 to WL1 + 400.
* Setting range
: 300 to 900 nm
* Initial (default) value : 700
* Pressing
U
key prior to the ENT key will return to
WL1.
(e) A memorize spectrum setting screen will now appear.
<Memorize spectrum setting screen>
MEMORIZE SPECTRUM NO. (1 - 4)
1
Set a memorize spectrum No. for the spectral data.
Press a numeric key 1 to 4 and then the ENT key.
* Initial (default) value: blank
* Upon entering a memorize spectrum No., the
previous spectral data will be deleted and the new
data will be memorized.
(f) The scan monitor screen appears, and spectral
acquisition within the set range is carried out.
<Scan monitor screen (spectral acquisition under way)>
SCANNING SPECTRUM NO.: 1 MODE: FL (Em)
Ex: 250 nm Em: *** nm
3 - 40
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...