3 - 61
3.5 Setup
3.5.1 Pump Mode Setting
This section shows how to set the pump in the single-liquid
delivery mode (FLOW mode).
NOTE
: Once this setting is made, it is memorized and need not
be made each time.
(to be continued)
SETUP: PUMP MODE CONTRAST S.NO
1 FLOW 4 02AB-111
SETUP ITEMS (PUMP=1 CONTRAST=2 D.MIX=3)
1
ENT
PUMP (1=1, 2=2, 3=3)
1
Select PUMP
1 ENT
SET UP
UTLTY
POWER ON +
Select PUMP number
3 ENT
GR MODE (LOW=1, FLOW=2)
2
Select FLOW (isocratic mode)
2 ENT
LOW :Low gradient mode
FLOW :Isocratic mode
Press the POWER and SET UP UTLTY keys simultaneously.
The Pump mode setting screen will be displayed.
Continue pressing SET UP UTLTY until the setting screen appears.
SETUP: PUMP MODE CONTRAST S.NO
3 FLOW 4 02AB-111
1 ENT
NOTE
: Only FLOW (isocratic mode) is selectable for the
PUMP3 setting.
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...