(13) Input the following contents as instrument conditions.
RACK CODE:1
SYRINGE SPEED:1
RACK PARAM:RACK CODE = 1
X1 = 57.1
X2 = 178.6
Nx = 10
Y1 = 0.0
Y2 = 148.5
Ny = 12
Z = 34
P = 5
INJ.METHOD: CUT
LEAD VOLUME: 30.0
REAR VOLUME: 30.0
ASP SYRINGE SPEED: 3
DSP SYRINGE SPEED: 2
AIR VOLUME: 2.0
NEEDLE WASH TIME SOLVENT1: 4
RINSE PORT WASH TIME: 1
SYRINGE VOL:2 (175 µL)
(14) Set the pump flow rate to 1.0 mL/min.
(15) Set the detector wavelength to 265 nm. Set the detector
time constant to 0.1 s.
(16) Set the following conditions in the data processing unit.
SAMPLE PERIOD : 100 ms
STOP TIME
: 3 min
4.1.3 Reproducibility
Check
Check the reproducibility of the injection volume in the following
procedure.
(1) Press
the
PROG
key and set the following conditions.
STEP RACK VIALS VOL(µL) INJ TIME
1 1 1- 1 10.0 6 3.0
Pressing the
ESC
key can complete setting.
(2) Prepare 60 % methanol for the mobile phase and wash
fluid. Wait until the pump, detector, etc. are warmed up
and the baseline stabilizes.
4 - 5
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...