3 - 19
3.2.2 Pump Motor Control Setting
The automatic flow rate control system (STANDARD) and the
fixed flow rate control system (FIX) are available as the pump
motor control method.
NOTE
: STANDARD is selected for ordinary use.
UTLITY (CONDITION=1, UNIT=2, FLUSH=3
1 EVENT=4, LINK=5, P.WASH=6, OTHERS=7)
CONDITION SET (Z-ADJ=1, MOTOR-CTRL=2
2 PURGE=3, CALIBRATION=4, LPGR=5)
Select CONDITION
1 ENT
MOTOR CONTROL (STANDARD=1, FIX=2)
1
Select MOTOR-CTRL
2 ENT
SET UP
UTLTY
Select STANDARD
1 ENT
Select FIX
2 ENT
The motor control method is set as STANDARD.
FIX STEP AREA (0, 30 - 300)
100
Enter FIX STEP AREA .
Input range is from 0 or 30 to 300.
ENT
Monitor screen
Monitor screen
Monitor screen
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...