1 ENT
USE ERROR OUT (YES=1, NO=0)
0
The ALL END signal will be output in ERROR OUT
on the e-Line. When ALL END “YES” is set, the
ALL END signal is output to the option input/output
circuit board irrespective of this setting.
For output in ERROR OUT, select “1” (YES) or “0”
(NO) for avoiding the output.
A wait time till output of the ALL END signal is
settable. For stopping the system with a short time
lag due to washing in the pump, a time period
should be set here taking a necessary time lag into
account.
15 ENT
Monitor screen reappears.
3 ENT
Select an operating condition.
1 ENT
For creating a new sequential program, input the
default rack code value at the first step.
1 ENT
OPERATION (RACK SELECT=1, INJ.METHOD=2,
1 APL PARAM=3)
RACK SELECT (1-16)
1
Monitor screen reappears.
ALL END TIME (0.1-100.0) min
15.0
(to be continued)
3 - 19
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...