2.5
2.5 Specifications
(1) Optical
system : Double-beam ratio photometric system
(dispersion with diffraction grating)
(2) Light source
: D2 lamp
W lamp
Hg lamp (for checking monochromator
wavelength)
(3) Measuring wavelength range
(a) 1-wavelength
measurement : 190 to 900 nm
In the D2&W mode, the D2 lamp is
used at 400 nm or shorter and the
W lamp is used at 401 nm or longer.
190 to 900 nm (D2&W mode)
190 to 600 nm (D2 mode)
380 to 900 nm (W mode)
(b) 2-wavelength
measurement : 2 wavelengths in wavelength
regions 190 to 350 nm,
351 to 400 nm, 401 to 600 nm and
601 to 900 nm (D2&W mode)
2 wavelengths in wavelength
regions 190 to 350 nm and 351 to
600 nm (D2 mode)
2 wavelengths in wavelength
regions 380 to 600 nm and 601 to
900 nm (W mode)
(Minimum wavelength interval 5 nm,
max. wavelength interval 160 nm
with data sampling period set at
400 ms)
(4) Spectral bandwidth
: 6 nm
(5) Wavelength
accuracy :
±
1 nm
(6) Time constant
: Selectable in 7 steps of 0.01,
0.02, 0.05, 0.1, 0.5, 1.0 and
2.0 sec
2 - 7
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...