(2) Connect the provided hypodermic syringe to the port4.
Hypodermic syringe
Fig. 5-7 Connecting the Hypodermic Syringe (or Tube)
(3) Using the SYR command, aspirate 175 µL of wash fluid.
FUNC
4
ENTER
1
7
5
ENTER
4
ENTER
(Refer to 3.2.2 Analyzing Operation (5) Functions (d) SYR
for details.)
NOTE
: When 70 µL optional syringe is used, using the
SYR command, aspirate 70 µL of wash fluid.
(4) Using the SVALVE command, select port 1 (RINSE PORT).
FUNC
6
ENTER
4
ENTER
(Refer to 3.2.2. Analyzing Operation (5) Functions (f)
SVALVE.)
(5) Loosen the plunger setscrew with your hand to remove the
syringe from the hook.
(6) Turn the top end (threaded metal part) of the syringe
counterclockwise with your hand to separate the syringe
from the syringe valve. Pull the plunger frontward out of
the hook.
5 - 9
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...