(a) Usual Grounding Connection (with grounded 3P power
outlet)
Grounded 3P power
outlet (15 A)
Grounded 3P
table tap
Power cord
(attached to
instrument)
Chromaster 5310
column oven
Ground resistance
100
Ω
or less
750 mm or
more below
ground
Connector
Rear panel
Copper plate
Grounding electrode
(underground section)
300 mm
Grounding construction f
100
Ω
or less
or
2 mm
10 mm
(Example of grounding
electrode)
Appended Fig. 1-1 Grounding Method
APPENDIX 1 - 5