(ii) Connection when grounded power outlet is
unavailable
Grounded 3P
table tap
Connector
Grounding wire (green)
Screw diameter :
M4 or larger
Turn count
:
3 or more
Wire thickness
:
1.25 mm
Grounding
terminal
Grounding construction for
100
Ω
or less
Grounding
resistance
100
Ω
or less
750 mm
or more
below
ground
Copper plate
300 mm
2 mm
10 mm
Chromaster 5210
autosampler
Rear panel
Grounding electrode
(underground section)
Grounded 2P power
outlet (15 A)
Grounded 3P
table tap
Power cord
(attached to
instrument)
Connecto
Grounding wire (green)
Screw diameter :
M4 or larger
Turn count
:
3 or more
Wire thickness
:
1.25 mm
Grounding
terminal
Grounding construction for
100
Ω
or less
Ground resistance
100
Ω
or less
Grounded 2P power
outlet (15 A)
Grounding electrode
(underground section)
Organizer or
r
AC adapter (option)
Thermostat
(ii-1) With 2P power outlet
APPENDIX 1 - 9
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...