3 - 14
SET PROGRAM: No. ( 1 – 9 )
1
EDIT MODE (MODIFY = 0, NEW = 1)
0
Select a program number from 1 to 9.
ENT
PRESSURE (MAX MIN)
(0.0 – 40.0) 20.0 0.0 1
Select EDIT MODE
ENT
Press the key to change the mode from manual set
screen to the programming mode.
Program generation is started
1
INITIAL
PROG
TIME FLOW EVENT
0.0 1.000 12 22 32 42 1
TIME FLOW EVENT
10.0 0.500 1
Input TIME, FLOW rate, and EVENT.
Up to 4 entries can be made per step for event.
Refer to 3.2.8 for details of EVENT.
ENT
TIME FLOW EVENT
1
A new input screen will be appeared.
Input TIME, FLOW rate, and EVENT.
ENT
Enter MAX and MIN pressure limit.
ENT
A new input screen will be appeared. Press the ESC key to
close the input screen, and return to the monitor screen.
ESC
Monitor screen
Monitor screen
INSERT
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...