(g) SPURGE
Syringe purging can be performed. The wash fluid a
syringe aspirated is discharged to the port4 side of the
syringe valve.
NOTE
:
Prior to pump washing, be sure to disconnect the stop
valve from port 7 of the syringe valve.
STEP RACK VIAL VOL(µL) INJ TIME [I]
1 1 1 10.0 1/1 0.0
FUNC
FUNC (TUBE=1, MOVE=2, HEIGHT=3, SYR=4,
7 VALVE=5, SVALVE=6, SPURGE=7, WPUMP=8)
7 ENT
ESC
MOVING [B]
IDLE WAIT!
3 ENT
SYRINGE PURGE (STROKES SPEED)
<1-20> 3 5
PRESS FUNC KEY OR IDLE KEY [W]
IDLE
FUNC
ESC
SYRINGE PURGE (STROKES SPEED)
<1-5> 3 5
5 ENT
ESC
MOVING [B]
IDLE WAIT!
Monitor screen reappears.
3 - 49