(2) Replacement of the Xe Lamp
WARNING
Electric Shock due to Contact with Inside of
Instrument
The xenon lamp is supplied with high voltage of 30 kV,
which could cause an electric shock to result in serious or
fatal injury.
Before removing the light source cover for replacement of
the light source lamp, be sure to turn off the power switch.
WARNING
Injury due to Xenon Lamp Explosion
If a strong shock or impact is applied to the xenon lamp or
if the surface of its quartz glass part is scratched, it may
explode and scatter glass pieces, resulting in personal
injury. Be sure to wear proper protective gear such as
safety goggles, safety mask, thick long sleeves, and
gloves when handling the xenon lamp.
(a) Turn OFF the power switch and unplug the power cord.
(b) Loosen six retaining screws of the light source cover,
and detach the cover (see Fig. 5-1).
Retaining screws
Fig. 5-1
Removal of Light Source Cover
5 - 5
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...