Fig. 2-8 All Volume Injection Method
Full Loop Injection Method
When applying Full loop injection method, the sample
introduction volume more than quintuple loop capacity
(injection volume) is required (amount of sample
necessity).
Select sample introduction volume within the area
indicated in the following figure.
The sample introduction volume is selected within the
shadow area of the following figure to fill a loop with the
density of the 99% of the sample. (For example, more
than 25 µL sample is required when injecting 5 µL
sample (5 µL loop capacity). This means that quintuple
sample volume is required.)
5
10
15
20
25
50
75
100
Amount of sample necessity introduction [µl]
S
am
ple I
nj
ect
io
n V
o
lum
e
[
µl]
(L
oop
Cap
acit
y)
Fig. 2-9 Full Loop Injection Method
2 - 11
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...