4.2.2 Wavelength accuracy check by using Hg lamp
Check the wavelength accuracy by using the bright line at 254
nm of the Hg lamp (see section 3.3.6). “GOOD” will appear on
the display when the wavelength error is within
±
3 nm, or “FAIL”
will appear when the error exceeds
±
3 nm.
NOTE
: Wavelength calibration is necessary when “FAIL”
appears. See section 4.2.4 (2) for the calibration.
4.2.3 Sensitivity
check
The S/N ratio of the Raman spectrum of water is used to
measure the sensitivity of the fluorescence detector.
The following items must be prepared for the measurement.
The following items are required.
•
Pump
•
Recorder or D-2500 chromato-data processor
•
Distilled water
(1) Injection of Distilled Water
Connect tubing as shown in Fig. 4-1, supply distilled water
at a flow rate of 1 mL/min for 3 minutes with the Chromaster
5110 Pump in Manual mode, then set the flow rate to 0 and
stop the supply of water. Remove the flow cell unit and
check for any bubbles in the flow cell.
NOTE
: If the flow cell has been left empty for awhile (including
when the instrument is installed), first inject methanol
or acetonitrile into the flow cell in order to wet the
interior in advance, then inject distilled water.
Injection of water without first injecting an organic
solvent may cause a reduction of sensitivity.
4 - 2
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...