2.3.1
2.3 Operating Principle and Functional Description
2.3.1 Temperature
Control
With the Chromaster 5310 column oven, analysis can be started
when oven temperature is within a certain set range
(temperature ready status). Parameters settable are Tolerance
and Wait Time, and when the measured temperature of the oven
is maintained within the set temperature
±
tolerance during the
wait time, the temperature ready status is assumed.
(See Fig. 2-6.)
Temperature
(
°
C)
Time (min)
Set
temperature
Wait time
Set temperature
±
tolerance
Temperature
busy status
Temperature
ready status
Fig. 2-6 Temperature Control
2 - 7