(b) When the time program settings are completed, the 1
st
monitor screen reappears.
(7) Addition of steps
Press the INSERT key to add a step to the program during
step input. The step setting screen will appear, so make
the addition in the same procedure as described above.
However, the addition must be made within the time
entered at the final step.
3.4.2 Time program starting
Press the START key to start the time program.
Status indication on the 1
st
monitor screen becomes R (time
program is running), and the elapsed time is indicated at TIME.
The PROG. RUN LED lights up in blue.
<1st monitor screen>
TIME Ex Em FL PROG LAMP [R]
*** 250 350 0.0
1
Xe
3.4.3 Time program stopping
Press the STOP key, then the time program under execution
stops and the 1
st
monitor screen returns.
3.4.4 Time program operation examples
(1) Setting of time program
Described next is the procedure for newly creating a time
program shown in the figure.
3 - 33
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...