<Utility setting screen>
UTILITY (OFFSET=1, TIME CONSTANT=2,
4
USE PROG=3, PMT VOLT=4, OTHERS=5)
Press numeric key 4 and then the ENT key.
(v) The pmt voltage setting screen appears.
<Pmt voltage setting screen>
PMT VOLTAGE (SUPER HIGH=1, HIGH=2, MID=3,
3
LOW=4, SUPER LOW=5)
Press numeric key 3 and then the ENT key.
(vi) The
1
st
monitor screen reappears.
(vii) Press
UTLTY
SET UP
key, and the utility setting screen
reappears.
<Utility setting screen>
UTILITY (OFFSET=1, TIME CONSTANT=2,
5
USE PROG=3, PMT VOLT=4, OTHERS=5)
Press numeric key 5 and then the ENT key.
(viii) The others setting screen appears.
<Others setting screen>
OTHERS (Em BANDWIDTH=1, ERROR OUT=2,
1
LEAK SENSOR=3, LAMP=4)
Press numeric key 1 and then the ENT key.
(ix) The Em bandwidth setting screen appears.
<Em bandwidth setting screen>
Em BANDWIDTH (STANDARD=1, WIDE=2)
1
Press numeric key 1 or 2 and then the ENT key.
*1: STANDARD = emission spectrum bandwidth 15 nm
2: WIDE = emission spectrum bandwidth 30 nm
4 - 5
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...