(b) Rack
parameter
Set the shape of the applied rack as a parameter.
(i) Contents of rack parameters
Rack Parameter
X1 : X coordinate of first vial within
55.0 to 180.0 mm
Y1 : Y coordinate of first vial within
0.0 to 155.0 mm
X2 : X coordinate of last vial within
55.0 to 180.0 mm
Y2 : Y coordinate of last vial within
0.0 to 155.0 mm
Z : Height of needle within
0.0 to 53.0 mm
Nx : Number of vials in X direction
1 to 40
Ny : Number of vials in Y direction
1 to 40
PATTERN CODE: Pattern code
Wash port
(29.7, 0.0, 53.0)
Idle position
Injection port
(0.0, 0.0, 40.0)
PI Origin position
X1
X2
Y2
Y1
x
Needle exchange
position
(57.1, 0.0, 10.0)
1 to 8
y
Fig. 3-5
Contents of Rack Parameters
Table 3-1
Rack Parameters
Rack
code X1 X2 Nx Y1 Y2 Z Ny
Pattern
code
Application
1 57.1
178.6
10 0.0
148.5
34.0
12 5
For 1.5ml
×
120 PCS (STD)
2 57.1
178.6
10 0.0
148.5
34.0
12 5
For 1.5ml
×
120 PCS (STD)
3 57.1
178.6
10 0.0
148.5
34.0
12 5
For 1.5ml
×
120 PCS (STD)
4 57.1
178.6
10 0.0
148.5
34.0
12 5
For 1.5ml
×
120 PCS (STD)
5 57.1
178.6
10 0.0
148.5
34.0
12 5
For 1.5ml
×
120 PCS (STD)
6 57.1
178.6
10 0.0
148.5
34.0
12 5
For 1.5ml
×
120 PCS (STD)
7 57.1
178.6
10 0.0
148.5
34.0
12 5
For 1.5ml
×
120 PCS (STD)
8 57.1
178.6
10 0.0
148.5
34.0
12 5
For 1.5ml
×
120 PCS (STD)
9 57.1
178.6
10 0.0
148.5
34.0
12 5
For 1.5ml
×
120 PCS (STD)
10 57.1 178.6
10 0.0 148.5
34.0
12 5 For 1.5ml
×
120 PCS (STD)
11 59.0 176.6
8 15.4
149.8
48.0
9
5 For 4ml
×
72 PCS
12 57.2 178.5
13 8.6 150.0
44.0
15 5 For 1ml
×
195 PCS
13 68.5 167.5
12 89.1
152.1
42.0
8
5 Back side of 96-holes microplate
14 68.5 167.5
12 2.8 65.8
42.0
8
5 Front side of 96-holes microplate
15 66.1 169.6
24 87.0
154.5
42.0
16 5 Back side of 384-holes microplate
16 66.1 169.6
24 0.5 68.0
42.0
16 5 Front side of 384-holes microplate
3 - 24
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...