3.4.7
Explained here is how to create a new time program
exemplifying the one shown above.
1) At start, the measuring wavelength is set at 250 nm
and auto zero adjustment is carried out for baseline.
2) After 10 minutes, the measuring wavelength is
changed to 270 nm. The baseline holds the previous
status (hold status).
3) After 20 minutes, the time program ends.
Press key.
Press [numerical] key [1], and then the [ENT] key.
Press [numerical] key [1], and then the [ENT] key.
Press [numerical] key [0], and then the [ENT] key.
Press [numerical] keys [2][5][0], and then the [ENT] key.
Press [numerical] key [1].
Press the [ENT] key.
Press [numerical] keys [1][0], and then the [ENT] key.
Press [numerical] keys [2][7][0], and then the [ENT] key.
Press numerical] key [2].
(Continued)
PROG
SET PROGRAM NO.
(
1–9
)
0
TIME
WL
BASE
(
A/Z=1, HOLD=2
)
_
1
TIME
WL
BASE (A/Z=1, HOLD=2)
0.0
250
1 1
TIME
WL
BASE (A/Z=1, HOLD=2)
0.0 250 1
1
EDIT MODE (MODIFY=0, NEW=1)
1
3 - 41
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...