(c) Put on safety goggles for the subsequent steps.
(d) Detach the connector between lamp house and main
body, and loosen the lamp house retaining screw.
Lamp house
Retaining screw
Grip
Connecter
Fig. 5-2
Removal of Connector and Retaining Screw
(e) Lift the grip of the lamp house lightly and pull it toward
you, then detach the lamp house from the main body.
Lamp house
Grip
Fig. 5-3 Removal of Lamp House
5 - 6