Appendix 1.3 Items to be Prepared by Customer
Appendix 1.3.1 Power Supply Arrangement
Confirm that the following requirements are satisfied before
installing the instrument.
(1) Power
Supply
Power supply
voltage
: 100 to 240 V AC
Fluctuation shall be within
±
10% of the
rated voltage.
Frequency
: 50 or 60 Hz
Fluctuation shall be within ±0.5 Hz of the
rated frequency.
Power capacity : 100 VA or more for a single instrument,
and 600 VA or more for the total system
The total system shown comprises
Chromaster organizer, Chromaster 5110
pump, Chromaster 5210 autosampler,
Chromaster 5310 column oven, and
Chromaster 5410/5420/5430/5450
detector.
Check the necessary power supply
requiretemnts and prepare a power
supply meeting the specified
requirements taking into account your
actual system configuration and the OA
equipment to be used in combination.
Power outlet
: Power supply shall be obtainable near
the place for installing the system
inclusive of this instrument.
This detector requires at least one
grounded power outlet (3-pole type).
At least one additional outlet is needed
for the entire system. (See section
2.5.2.)
APPENDIX 1 - 4
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...