(3) With the tube set in the pre-heating base, re-attach the pre-
heating plate and fasten it with the two hexagon bolts.
NOTE
: Before re-attaching the pre-heating plate, be sure
to make sure that the tube is neatly installed in the
grooves in the pre-heating base.
Re-attaching the pre-heating base with the tube
not properly set may damage the tubing or lead to
incorrect results.
5.4 Assembling of Pre-heating Tubing (for Pre-heating of 0.5 Line or of More Than
1.5 Line)
Where a flow rate at lower than 0.7 mL/min is used, set the
tubing from the center of the pre-heating section as shown in
Appended Fig. 1-11 below (0.5 line).
Column connected
side
100 mm
Set the tubing from
the center.
Appended Fig. 1-11 Assembling of Tubing (for Pre-heating
of 0.5 Line)
APPENDIX 1 - 15
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...