(6) Setup
Processing
You can select use or nonuse of the thermostat (option),
and set LCD contrast of the UI pad.
(a) Selection of use or nonuse of thermostat (option)
ENT
SETUP: THERMO UNIT CONTRAST S.NO.(A/S)
USE 4 02AB-111
1 ENT
THERMO UNIT (NO USE=0, USE=1
)
1
1 ENT
SETUP ITEMS (THERMO UNIT=1, CONTRAST =2)
1
ENT
Turn off power supply
and turn it on again.
UTLTY
(
SET UP
)
POWER ON
SETUP: THERMO UNIT CONTRAST S.NO.(A/S)
NO USE 4 02AB-111
+
Turn on power supply with the
UTLTY
key held down.
Indicates a serial No. (product No.).
Enter 0 for avoiding use of the thermostat, and 1 for
selecting its use.
When turning on power supply again, the instrument starts
under the set conditions.
3 - 51
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...