(2) Wash fluid selection and use examples
The Chromaster 5210 autosampler uses two kinds of wash
fluid. Refer to the following table.
Table 2-2
Isocratic Analysis
One Fluid Wash <standard>
(for Cut injection method,
Full loop injection method)
Two Fluids Wash
(for All volume injection method,
Low carry over analysis)
Wash fluid 1
Sample solution
*1)
,
Mobile phase
*1)
Sample solution
*1)
(Strong wash)
Wash fluid 2
Aqueous solvent
*2)
Mobile
phase
*1)
(Weak wash)
Table 2-3
Gradient Analysis
One Fluid Wash <standard>
(for Cut injection method,
Full loop injection method)
Two Fluid Wash
(for All volume injection method,
Low carry over analysis)
Wash fluid 1
Sample solution
*1)
,
Sample solution
*1)
Gradient final solvent
*3)
(Strong wash)
Wash fluid 2
Aqueous solvent
*2)
Gradient initial solvent
*1)
(Weak wash)
*1) As a washing solution in cases where a solvent
containing a buffer solution or salt has been fed as the
mobile phase, use a mobile-phase solvent containing no
salt.
(When 20 mmol/L KH2PO4:CH3CN
=
50:50 is used for
the mobile phase, use H2O:CH3CN
=
50:50.)
If a solvent containing salt is used for washing, it causes
damages, such as the surface of the glass, the tip of the
syringe, and the inside of the wash pump.
Additionally, it is resulting in deterioration of performance
or shorter service life.
*2) When the aqueous solvent is used for washing of a
plunger seal for the Chromaster 5110 pump.
For example, when an organic solvent is used as wash
fluid1 in the normal phase chromatography analysis
which the cut injection method is applied, salt
precipitation in the pump plunger can’t be washed off
(Salt doesn’t dissolve). In that case, use the aqueous
solvent as wash fluid 2.
2 - 12
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...