3.3.5 Checking the Wavelength Accuracy with Hg Lamp
It is checked whether the wavelength is normal or not using the
254 nm bright line of mercury lamp.
1)
Press key.
2)
The GLP setup screen appears.
GLP Setup Screen
Select Hg CHECK.
Press [numerical] key [4], and then the [ENT] key.
3)
Checking action will automatically be taken.
During check, the following display is presented.
GLP
CHECKING WL ACCURACY
(Hg)
GLP (LOGBOOK=1, KEY LOCK=2,
1
D2 CHECK=3, W CHECK=4, Hg CHECK=5)
4)
After this check, its result is indicated.
ACCURACY (254 nm)
GOOD
-1nm
Wavelength accuracy is checked utilizing the bright line of
the mercury lamp at 254 nm. When deviation is within
±
1
nm, “GOOD” will be indicated. If deviation exceeds
±
1 nm,
“FAIL” will be indicated. When no bright line has been
detected, “NOT DETECTED” is indicated. In this case,
confirm mounting of the flow cell, and then turn on power
supply again. If the situation does not improve, contact the
nearest maintenance service company authorized by
Hitachi High-Technologies.
3 - 29
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...