3.4 Analyzing Operation
(b) Storing the Background Spectral Data in Memory (“2” key
followed by ENT key)
Scan is automatically carried out.
During acquisition of the background spectrum, the
following screen is presented.
Upon completion of spectral data acquisition, the monitor
screen automatically appears.
(c) Storing the Reference Energy with D
2
lamp in Memory (“3”
key followed by ENT key)
Set the minimum wavelength of scan range.
Press [numerical] keys (within [2][0][0] to [8][5][0]), and then
the [ENT] key.
* Enterable range: 200 to 850 nm, enterable in 1-nm
steps
Set the maximum wavelength of scan range.
Press [numerical] keys (within [2][5][0] to [9][0][0]), and then
the [ENT] key.
* Enterable range: Minimum scan range is 50 nm.
Measuring span is max. 400 nm.
Input the number of spectral data.
WL: *** nm
SCANNING WL RANGE (WL1 WL2) nm D2 REF
<200 – 850>
200 – 600
SCANNING WL RANGE (WL1 WL2) nm D2 REF
<250 – 600>
200 – 600
MEMORIZE SPECTRUM NO. (1–3)
_
SCANNING BACK GROUND SPECTRUM
3 - 32
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...