5.2.1
5.2 Parts Replacement Method
5.2.1 Fuse
Replacement
WARNING
Electric Shock due to Contact with Inside of
Instrument
When replacing the fuse with a new one, be careful not to
receive an electric shock. To prevent this, be sure to turn
off the power switch before proceeding to fuse replacement.
(1) Item to Prepare
New fuse (P/N J821349)
Make sure to use a time lag fuse of the specified rating
(T5AL for 250 V).
(2) Check
Items
Make sure the power switch is OFF and the power cord is
disconnected from the instrument prior to fuse replacement.
(3) Fuse
Replacement
(a) Detach the fuse holder by pressing the knobs at both
sides.
Knobs
C941322
Fuse holder
Fig. 5-1 Fuse Replacement
(b) Replace the fuse and then re-attach the fuse holder by
pressing it.
5 - 3
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...