Full Loop Injection Method Procedure
The full loop injection method, the injection volume, the west
volume, and the air volume are restricted by the following relational
expression. A stroke of the syringe capacity can’t be used all.
<175 µL of Standard syringe >
vi + vw + va
≦
165 µL
vi :Injection volume (The sample net volume to the flow path)
vw :Waste volume (The disposal sample volume of the lead and rear
part of the loop)
va :Air volume (The air volume of the lead and rear part of a sample)
One-half of the waste volume (vw/2) is calculated and delivered
automatically to the lead and the rear part of the loop each.