5.12 How to Replace Fuse (in the optional thermostat)
WARNING
Electric Shock due to Contact with Inside of Instrument
When replacing the fuse with a new one, be careful not to
receive an electric shock. To prevent this, be sure to turn off
the power switch before proceeding to fuse replacement.
(1) Item to be Prepared
New fuse (P/N J821336)
Be sure to use a new time-lag fuse meeting the specified
(2) Pre-check
Before fuse replacement, make sure that the power switch
of the instrument is off and that the power cord is
disconnected from it.
(3) Fuse Replacement Procedure
(a) Remove the fuse holder while holding the tabs on both
sides.
Tabs
Fuse holder
C941322
(b) Replace the existing fuse with a new one, and insert
the fuse holder in place.
5 - 25
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...