APPENDIX 2 - 3
3. Individual
Contacts of Each Unit
(1) Contact Signal Input Terminals
These are contact signal input terminals for control using
contact signals from an external switch, relay etc. Any of
the signals is activated by shorting the terminal for 0.1
second or longer. The input circuit is shown in Appended
Fig. 2-3.
(2) Contact signal Output Terminals
These are contact signal input terminals for control using
contact signals of an external unit. The output circuit is
shown in Appended Fig. 2-4. The rating of contacts used
is 12 V,
0.1 A. Make sure the load connected does not exceed
this rating.
Appended Fig. 2-1 e-Line Cable Connection
Appended Fig. 2-2 e-Line Contact Circuit Configuration
e-Line cable
Other unit
Chromaster series unit A
Chromaster series unit B
+5 V
320
IN
Max. 10 mA/unit
Max. 0.1 A
OUT 4.7 k
ON : 5 V
OFF :
0 V
Max. 12 V
(reverse polarity
not allowed)
(H)
(L)
(G)
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...