3.2.5
(c) Setting the Use (ON) or Non-use (OFF) Status of Leak
Sensor
1) Press key.
SET UP
UTLTY
2) The utility setup screen appears.
Utility Setup Screen
Select OTHERS.
Press [numerical] key [5], and then the [ENT] key.
3) The OTHERS setup screen appears.
OTHERS (PROCESSOR=1, ERROR OUT=2
1
LEAK SENSOR=3, LAMP=4, GAIN=5)
UTILITY (OFFSET=1, TIME CONSTANT=2,
1
USE PROG=3, POLARITY=4, OTHERS=5)
Select LEAK SENSOR.
Press [numerical] key [3], and then the [ENT] key.
4) The LEAK SENSOR setup screen appears.
1
LEAK SENSOR (OFF=0, ON=1)
Set the use (ON) or non-use (OFF) status of LEAK
SENSOR.
Press [numerical] key ([1] or [0]), and then the [ENT]
key.
* Default value: 1
5) Upon completion of LEAK SENSOR setting, the first
monitor screen reappears.
NOTE
: If power supply is turned off with the leak sensor
set at OFF, you should remember that the “ON”
setting returns upon turning on power supply
again.
3 - 17
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...