If emission occurs only in the entrance area for excitation
beam, such a sample needs to be diluted with an
appropriate factor before measurement.
The second factor is called “concentration quenching”
where activity is disturbed by interaction of molecules.
Another potential error that may occur when fluorescence
detection is used is the re-absorption of fluorescence (self-
absorption of fluorescence). An example of this
phenomenon is shown in Fig. 3-9. Re-absorption of
fluorescence occurs when the tail of the shorter-wavelength
side of fluorescence spectrum overlaps the tail of the long-
wavelength side of the excitation spectrum. When this
phenomenon occurs, the fluorescence spectrum appears to
be somewhat shifted toward the longer-wavelength.
This phenomenon will very rarely cause a substantial error
in quantitation when fluorescence detection is used, and
should not be a cause for concern.
Relati
ve intensity
Wavelength
Fluorescence re-absorbed
Fluorescence
spectrum
Excitation
spectrum
Fig. 2-9 Re-absorption of Fluorescence
Anyway if a high concentration may have caused a
measurement error, dilution with an appropriate factor is
necessary.
2 - 13
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...