3.2 UTLTY Setting
3.2 UTLTY Setting
3.2.1 Setting of Leak Sensor
(1) Set the leak detection level for the gas leak sensor.
UTILITY (LEAK SENSOR=1, ANALOG OUT=2,
1 ERROR OUT=3, CALIBRATION=4)
SELECT SENSOR (GAS=1, LIQUID=2)
1
LEAK GAS (MONITOR=1, SET LEVEL=2)
1
SENSOR MONITOR
ALARM LEVEL
400 1200
Monitor screen
SET ALARM LEVEL (0 - 4095)
1200
Monitor screen
1 ENT
1 ENT
Monitor the gas leak sensor
output. 1 ENT
2 ENT
SET UP
UTLTY
ENT or ESC key
Set up a gas leak detection level.
1200 ENT (initial value: 1200)
3 - 10