3 - 22
3.2.5 Pressure Calibration Setting
Set up a calibration coefficient for the pressure.
UTLITY (CONDITION=1, UNIT=2, FLUSH=3
1 EVENT=4, LINK=5, P.WASH=6, OTHERS=7)
CONDITION SET (Z-ADJ=1, MOTOR-CTRL=2
4 PURGE=3, CALIBRATION=4, LPGR=5)
Select CONDITION
1 ENT
CALIBRATION (FLOW=1, PRESS=2)
2
Select CALIBRATION
4 ENT
SET UP
UTLTY
Monitor screen
Select PRESS
2 ENT
PRESSURE CALIB (0.900 – 1.100)
1.000
Monitor screen
Enter calibration coefficient.
ENT