3 - 24
3.2.7 Column Flush Setting
The pump automatic stop function and the column flushing
function after the completion of analyses is set.
By setting of the column flush, the flushing will start
automatically at specified time after the analysis.
When a salt solution is used for analysis, the column must be
flushed with water to prevent elution of the salt component.
As an example, execution of the pump automatic stop function
or the column flushing function were set up after 15 minutes of
the analysis, and the column flushing time is set up as 30
minutes.
(to be continued)
UTLITY (CONDITION=1, UNIT=2, FLUSH=3
3 EVENT=4, LINK=5, P.WASH=6, OTHERS=7)
AUTO PUMP OFF (YES=1, NO=0)
1
Select FLUSH
3 ENT
SET DELAY TIME (1.0 – 600.0) min
15.0
Select YES
1 ENT
SET UP
UTLTY
Monitor screen
Enter DELAY TIME for execution of the pump auto-stop
function and the column flush function from 1.0 to 600.0 min.
1 5 ENT
COLUMN FLUSH (YES=1, NO=0)
1
Select YES
1 ENT
FLUSH MODE (PARAM=1, PROGRAM NO.9=2)
1
1=flushing by the specified condition
2=flushing by the gradient program
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...