4.4.3 Leakage from the flow cell
It is important to check for any liquid leakage from the flow cell
unit prior to analysis. If a leak is found, stop the pump
immediately and remove the flow cell unit. Take the following
procedure to remedy the leak.
Inlet tube
Setscrew 2
Setscrew 1
Outlet tube
Fig. 4-14 Inlet and Outlet Tube
(1) Supply liquid using the pump, and check that liquid comes
out of the outlet tube of the flow cell.
(2) If liquid does not come out of the tube, it indicates the tube
is clogged. Replace the outlet tube with a new one.
(3) If a leakage occurs though liquid comes out of the tube
normally, setscrew tightening may be slightly inadequate.
Retighten setscrew 1 lightly (about 10 to 20
°
) with a wrench
or setscrew 2 lightly (about 10 to 20
°
) manually.
(4) Wipe up any liquid spilt on the outside of the flow cell.
(5) Before returning the flow cell unit into position, supply liquid
with the pump and make sure there is no more leakage.
NOTE
: It is important that the OUTLET tube of the flow cell be
checked first for clogging. If the upper setscrew is
judged to be loose and is retightened with the wrench
without checking the OUTLET tube, then the flow cell
may be damaged or broken.
NOTE
: If leakage is found in the sample compartment, wipe up
the liquid thoroughly and then clean with a cloth
moistened with water.
4 - 29
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...