5.7 Replacing the Injection Valve Seal
NOTE
: Before proceeding to replacement of the injection valve
seal, be sure to stop the back flush pump.
Items to be prepared:
Hexagonal wrench (9/64 inch, This wrench is attached to P/N
893-0812)
Wrench (1/4 inch)
Injection valve seal (P/N 893-0812),
(1) Using a wrench (1/4 inch), disconnect the tubes and pipes
from the injection valve port 1 to 5.
Port 1
: from the pump
Port 2
: to the column
Port 5
: drain tube
Port 3 and 6 : Loop
Drain pipe
Loop
Injection valve
Screw cap
Fig. 5-15 Disconnecting Tubes and Pipes
from the Injection Valve
NOTE
: Avoid loosing the screw cap.
Otherwise, it may cause a reduction in service life
or performance.
(2) Loosen three stator setscrews using a hexagon wrench
(9/64 inch), and demount the injection valve stator.
(3) Pull out a stator ring.
5 - 16
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...