3 - 16
Fig. 3-3a Manual wash
(1) Set up a waste solution bottle (or beaker) under the drain
tube.
NOTE
: Please be careful about a tube tip position.
When the tube tip touches liquid level, a liquid lead
can’t be detected.
(2) Take wash fluid (distilled water) by a syringe of an
accessory.
(3) Connect the syringe to a joint of an inlet tube, and the wash
fluid (distilled water) is injected.
(4) Wash fluid is injected until a flow path of the wash system
is filled with it.
(5) Detach the syringe from the joint of the inlet tube and stop
up quickly.
(6) The waste solution bottle is emptied and rearranged.
Waste
solution bottle
(beaker)
Syringe
Inlet tube
Drain tube
Joint