(2) Cleaning of Liquid Leak Sensor
<Item to be prepared>
•
Hexagonal wrench (P/N: N315003, size: B2.5)
(a) Separate the liquid leak sensor from the drain tray by
removing the liquid leak sensor fastening screw with
the furnished hexagonal wrench.
(b) Wipe off the water adhering to the liquid leak sensor
and drain tray, and then fasten the sensor to the drain
tray again.
NOTE
: While cleaning the liquid leak sensor, be careful not to
damage its tip. If damaged, the liquid leak sensor may
not work properly or an error may occur.
Liquid leak sensor
fastening screw
Fig. 4-7 Liquid Leak Sensor Fastening Screw
NOTE
: Mount the leak sensor at the correct position. If the
sensor is mounted at a wrong position, it may not work
properly or an error may occur.
4 - 19
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...