3.2 Utility Setting
Set the full scale of processor output.
Press [numerical] key (within [1] to [4]), and then the
[ENT] key.
* Default value: 3
5) Upon completion of processor setting, the first monitor
screen reappears.
(b) Setting the Enable or Disable Status of Error Signal Output
SET UP
1) Press key.
UTLTY
2) The utility setup screen appears.
Utility Setup Screen
1
USE PROG=3, POLARITY=4, OTHERS=5)
UTILITY (OFFSET=1, TIME CONSTANT=2,
Select OTHERS.
Press [numerical] key [5], and then the [ENT] key.
3) The OTHERS setup screen appears.
1
LEAK SENSOR=3, LAMP=4, GAIN=5)
OTHERS (PROCESSOR=1, ERROR OUT=2
Select ERROR OUT.
Press [numerical] key [2], and then the [ENT] key.
4) The ERROR OUT setup screen appears.
1
ERROR OUT SIGNAL (DISABLE=0
,
ENABLE=1)
Set the ENABLE or DISABLE status of ERROR OUT.
Press [numerical] key ([1] or [0]), and then the [ENT]
key.
* Default value: 1
5) Upon completion of ERROR OUT setting, the first
monitor screen reappears.
3 - 16
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...