4 - 14
No. Symptom
Cause
Measure
7
Liquid doesn’t flow.
1) Looseness of the piping joint.
2) Clogging of the solvent filter.
3) Negative pressure depending on
the position of the solvent bottle.
4) Impurities or clogs in the flow
path.
1) Tighten the joint manually.
2) Wash the solvent filter.
3) Place the solvent bottle in
the same height as the
instrument.
4) Wash the piping, the mixer,
and the inside of the
degassing unit enough.
When using the solution
which is salt-induced
precipitation type, wash with
distilled water enough, and
replace with the target
solvent completely.
8
A leak error occurs
1) The liquid leaked.
2) Dirt attached to the leak sensor.
1) Clean up the leaked liquid
in the drain tray. Wipe the
liquid attached to the leak
sensor off. Readjust the
height of the leak sensor
(refer to 4.2.1).
2) Wipe the liquid attached to
the leak sensor off.
Readjust the height of the
leak sensor (refer to 4.2.1).
NOTE
: The whirring sound of the pump during operation varies
depending on such operating conditions as flow rate
and pressure. Although the whirring sound may
become larger anywhere in the high-pressure range, it
is not a symptom of abnormality.
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...