SAFETY - 3
SAFETY SUMMARY
WARNING precautions in the Manual
WARNING
Ignition of Flammable Chemicals
•
This instrument is not explosion-proof. In unattended operation, do not use organic
solvents having an ignition point below 70
°
C
(Section 3.6, 4.4)
•
Beware of ignition hazard when using flammable chemicals such as organic solvents.
(a)
Do not bring a heat or flame source near the instrument.
(b)
Well-ventilate the laboratory room where the instrument is used.
(c)
Always check the following conditions. If an abnormality is found, stop operation
immediately.
◊
Leakage of solvent or waste solution.
◊
Leakage of solvent inside the instrument.
(Section 3.6, 4.4)
•
When using flammable chemicals, be careful about possible ignition due to static
electricity. To prevent the build-up of static electricity, use a conductive container for
waste solution employ and provide proper grounding connection to it.
(Section 3.6, 4.4)
Explosion of Vapor from Flammable Chemicals
If a flammable chemical such as organic solvent leaks from the flow path of the instrument
and its vapor concentration reaches the explosion limit, it could cause spontaneous
combustion with dangerously explosive results.
When using a flammable and readily volatile chemical, be sure to check for leakage from
the instrument flow path and ventilate the laboratory room adequately.
(Section 3.6, 4.4)
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...