(4) Confidence-level
reporting
This autosampler incorporates the confidence-level
reporting function for analysis to be carried out while
checking data reliability.
For using this function, press the
GLP
key.
GLP
Input “LOGBOOK” for selecting a logbook which records
maintenance information or “KEY LOCK” for invalidating key
input in order to prevent an erroneous operation.
1 ENT
Input the number of the part to be checked or reset.
1 ENT
Select 1 for logbook display or 2 for resetting the
logbook.
1 ENT
This screen displays the number of needle
insertions into the injection port after the latest
replacement of the injection port seal and the
date of the replacement.
ENT
or
ESC
STEP RACK VIAL VOL(µL) INJ TIME [I]
1 1 1 10.0 1/1 0.0
GLP (LOGBOOK=1, KEY LOCK.=2)
1
LOGBOOK (REPORT=1, RESET=2)
1
Monitor screen reappears
INJ. PORT SEAL LAST CHANGED DATE
23 times 1/1/2000
PARTS (INJ. PORT=1, INJ. VALVE=2
1 SYR=3, SVALVE=4, WASHING PUMP=5)
(to be continued)
3 - 35
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...